| Specifications |
Size Range:
|
5 - 100μm |
Channel
Sizes:
|
Standard: 5, 10, 25, 40, 50, 100μm |
Counting
Efficiency:
|
50% @ 5.0μm; 100% for particles
> 7.5μm (per JIS) |
Flow Rate:
|
1.0 CFM (28.3 LPM) |
Laser Source:
|
Laser Diode |
Zero Count
Level:
|
<1 count / 5 minutes (per JIS) |
| Concentration Limits:
|
500,000 Particles/ft3 @5% Coincidence
Loss |
Calibration:
|
NIST Traceable |
Count Modes:
|
Automatic, Manual, Beep, Concentration,
Cumulative/Differential, Real-time
(with LMSNet, LMS Express RT/RT+) |
Data Storage:
|
3,000 Sample Records (Rotating
Buffer) which include particle and
environmental data, locations and
times |
Communication
Modes:
|
RS-232 via RJ-45 to
PC, RS-485/MODBUS |
Supporting
Software:
|
LMS XChange, LMS Express, LMS
Express RT and RT+; LMSNet |
| Environmental Sensors:
|
Optional (up to 4): Temperature/Relative
Humidity, Air Velocity, Differential
Pressure, or others |
Touch Screen
Display:
|
5.7” (14.47 cm) Color, 320x240
pixels |
| Printer: |
Built-in Thermal Printer |
| Reports: |
Normalized Counts, FS-209E, ISO-14644-1 |
| Alarm: |
Integrated Alarm, External Alarm
Connector; Alarms on Counts or Low
Battery |
| Enclosure: |
Stainless Steel |
| Sample Output: |
Internally Filtered to HEPA Standards
(>99.997% @ 0.3μm) |
Vacuum
Source:
|
Internal Pump, Automatic Flow
Control |
| Power: |
100-240 VAC, 50-60 Hz |
| Battery: |
Li-Ion, Removable & Rechargeable |
Dimensions:
|
9.96” (l) x 8.0” (w) x 10.16”
(h) [25.3 x 20.32 x 25.81 cm] |
| Weight:
|
With Battery: 14.9 lbs (6.8 kg)
Without Battery: 12.95 lbs (5.9
kg) |
| Environmental Conditions |
| Operating:
|
50°F to 104°F (10°C
to 40°C) / 20% to 95% non-condensing |
| Storage:
|
14°F to 122°F (-10°C
to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included:
|
Operating Manual on
CD; Isokinetic Sample Probe; Purge
Filter; Removable, Rechargeable
Li-Ion Battery; Power Cable and
Spare Fuse; LMS XChange Software;
Printer; Thermal Printer Paper |
| Optional:
|
Printed Operating Manual;
Network Adapter; Wireless Network
Adapter; Carrying Case with Wheels;
Cleanroom Cart; Temperature/Relative
Humidity Probe; Air Velocity Probe;
Differential Pressure Probe; High
Pressure Diffuser; Spare Li-Ion
Battery; External Battery Charger;
LMS Express, Express RT and RT+
Software; LMSNet Software |