| Specifications |
Size Range:
|
0.2 - 2.0μm |
Channel
Sizes:
|
Standard: 0.2, 0.3, 0.5, 0.7,
1.0, 2.0μm |
Counting
Efficiency:
|
50% @ 0.2μm; 100% for particles
> 0.3μm (per JIS) |
Flow Rate:
|
0.1 CFM (2.83 LPM) |
Laser Source:
|
Laser Diode |
Zero Count
Level:
|
<1 count / 5 minutes (per JIS)
|
| Concentration Limits:
|
4,000,000 Particles/ft3 @5% Coincidence
Loss |
Calibration:
|
NIST Traceable |
Count Modes:
|
Automatic, Manual, Beep, Concentration,
Cumulative/Differential and Real
Time (with LMSNet or LMS Express
RT/RT+) |
Data Storage:
|
3,000 Sample Records (Rotating
Buffer) which include particle and
environmental data, locations and
times |
Communication
Modes:
|
RS-232 via RJ-45 to
PC, RS-485/MODBUS |
Supporting
Software:
|
LMS XChange, LMS Express, Express
RT and RT+, LMSNet |
| Environmental Sensors:
|
Optional Sensors (up to 4) include
Temperature/Relative Humidity, Air
Velocity and Differential Pressure |
Touch Screen
Display:
|
5.7-inch (14.47 cm) Color, 320x240 |
| Printer: |
Built-in Thermal Printer |
| Reports: |
FS-209E, ISO-14644-1 and EU GMP |
| Alarm: |
Integrated, Adjustable Alarm,
External Alarm Connector. Alarms
on Counts, Low Battery, Sensor Failure |
| Enclosure: |
Stainless Steel |
| Sample Output: |
Internally Filtered to HEPA Standards
(>99.997% @ 0.3μm) |
Vacuum
Source:
|
Internal Pump, Automatic Flow
Control |
| Power: |
100-240 VAC, 50-60 Hz |
Battery:
|
Li-Ion, Removable & Rechargeable |
Dimensions:
|
8” (l) x 8.5” (w) x 8.9” (h) [20.32
x 21.59 x 22.60 cm] |
| Weight:
|
With Battery: 14.75
lbs (6.7 kg) Without Battery:
12.75 lbs (5.8 kg) |
| Environmental Conditions |
| Operating:
|
50°F to 104°F (10°C
to 40°C) / 20% to 95% non-condensing |
| Storage:
|
14°F to 122°F (-10°C
to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included:
|
Operating Manual on
CD; Isokinetic Sample Probe with
Tripod and Tubing; Purge Filter;
Removable / Rechargeable Li-Ion
Battery; Power Cable and Spare Fuse;
LMS XChange Software; Printer; Printer
Paper |
| Optional:
|
Printed Operating Manual;
Network Adapter; Wireless Network
Adapter; Carrying Case with Wheels;
Cleanroom Cart; Temperature/Relative
Humidity Probe; Air Velocity Probe;
Differential Pressure Probe; Filter
Scanning Probe; High Pressure Diffuser;
Spare Removable/Rechargeable Li-Ion
Battery; External Battery
Charger; LMS Express, LMS Express
RT or RT+ Software; LMSNet Software;
Validation Documentation |