| Specifications |
Size Range:
|
0.3 - 25.0μm |
Channel
Sizes:
|
0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm
0.3, 0.5, 1.0, 5.0, 10.0, 25.0μm |
8-Channel:
|
0.3, 0.5, 0.7, 1.0, 3.0, 5.0,
7.0, 10.0μm |
Flow Rate:
|
1.0 CFM (28.3 LPM) |
Counting
Efficiency:
|
3100Rx = 50% @ 0.3μm; 100%
for particles > 0.45μm Per
JIS B 9921, ISO 21501-4 |
Light Source:
|
Extreme Life Laser Diode Technology
with >20-year Mean Time To Failure
(MTTF) |
Zero Count
Level:
|
<1 count / 5 minutes (per JIS
B 9921) |
| Concentration Limits:
|
500,000 particles / ft3 @ 5% coincidence
loss |
Calibration:
|
NIST Traceable, per ISO 21501-4 |
Count Modes:
|
Manual, Automatic, Concentration
and Real Time (with LMSNet or LMS
Express RT/RT+) |
Data Storage:
|
3,000 Sample Records (Rotating
Buffer) which includes particle
data, locations and times |
Communication
Modes:
|
RS-485/MODBUS; Ethernet;
Universal Serial Bus |
Supporting
Software:
|
LMS XChange, LMS Express, LMS
Express RT/RT+, LMSNet; LMS-Pharma
Edition |
Touch Screen
Display:
|
5.7-inch (14.47 cm) color, 320x240 |
| Printer: |
Optional Built-in Thermal Printer |
| Reports: |
ISO-14644-1, FS-209E & EU
GMP Annex 1 |
| Alarm: |
Integrated, Adjustable Alarm:
Alarms on Counts and Low Battery |
Enclosure:
|
316L Stainless Steel |
| Sample Output: |
Internally filtered to HEPA standards
(>99.997% @ 0.3μm) |
Vacuum
Source:
|
Internal flow-controlled pump |
| Power: |
100-240 VAC, 50-60 Hz |
| Battery: |
Li-Ion, Removable & Rechargeable |
Dimensions:
|
9.75” (d) x 8.0” (w) x 10.25”
(h) [24.77 x 20.32 x 26.03 cm] |
| Weight:
|
With Battery: 16.6 lbs (6.8 kg);
Without Battery: 14.65 lbs (5.9
kg) |
| Environmental Conditions |
| Operating:
|
50°F to 104°F (10°C
to 40°C) / 20% to 95% non-condensing |
| Storage:
|
14°F to 122°F (-10°C
to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included:
|
Operating Manual on
CD; Data Cable; Isokinetic Sample
Probe with Tripod and Tubing; Purge
Filter; Removable/Rechargeable Li-Ion
Battery; Power Cable and Spare Fuse;
LMS XChange Software; Printer Paper
(with Printer Option) |
| Optional:
|
Printed Operating Manual;
Thermal Printer; Cleanroom Cart;
Spare Removable/Rechargeable Li-Ion
Battery; External Battery Charger;
LMS Express/RT/RT+ Software; LMSNet
Software; Validation Manual |