| Airborne Particle Counters |
| Portable Products
> SOLAIR 5010+ |
| SOLAIR 5010+ |
| |
| Solair
5010+ Portable Particle Counter
The SOLAIR 5010+
line of portable counters offers
a sensitivity of 0.3 micron and
a flow rate of 0.1 CFM (2.83 LPM)
with a wide dynamic range up to
25.0 microns.
Utilizing a large, color 5.7”
(14.47 cm) touch screen interface,
the SOLAIR 5010+
is easy to configure and operate.
The SOLAIR 5010+
can store a large amount of particle
count data from 6 channels and
data from up to 4 environmental
sensors (4-20 mA). All data can
be quickly and reliably downloaded
to a computer or printed to its
built-in thermal printer.
Demonstrating its versatility,
the SOLAIR 5010+
can be used as a portable instrument
or integrated with a larger facility
monitoring system.
Designed and built by Lighthouse
Worldwide Solutions,
a name you can trust. |

[ Virtual
Demo ] |
|
| |
 |
 |
 |
 |
 |
 |
| New Easy Loading Thermal Printer |
User Removable Rechargeable Li-Ion
Battery |
Zoom Screen
Functionality |
Alpha Numeric
Location Labels |
User Friendly
Configuration |
Multiple
Cleanroom
Standard Reports
via Printer |
|
|
| Features |
Benefits |
- 0.5 µm Sensitivity
- 0.1 CFM (2.83 LPM) Flow Rate
- View 6 Particle Sizes Simultaneously
- Wide Dynamic Range, 0.3 to
25.0 µm
- Meets JIS Standards
- 5.7” (14.47 cm) Color Touch
Screen Interface
- Removable/Rechargeable Li-Ion
Battery
- Large Memory for Storing Data
(3,000 samples)
- Up to 4 Environmental Sensors
(4-20 mA)
- Alphanumeric Location Labels
- Easy to Load Thermal Printer
- User Definable Printouts
- Data Reported in Normalized
Counts, FS-209E, ISO-14644-1,
BS-5295 or EC GMP
- Easily Configurable Interface
with Zoom Capability
- Internal Audible Alarm
- External Alarm Trigger
- Data Easily Downloaded to
Computer
|
- 2 Year Warranty
- User Friendly Interface
- Easy-to-Read Display
- International Sales and Service
Support
- Low Cost of Ownership
|
|
|
| |
| Specifications |
Size Range:
|
0.5 - 25.0 µm |
Channel
Sizes:
|
Standard: 0.5, 0.7, 1.0, 3.0,
5.0, 10.0 µm
Standard: 0.5, 1.0, 3.0, 5.0,10.0,
25.0 µm |
Counting
Efficiency:
|
50% @ 0.5µm; 100% for particles
> 0.75µm (per JIS) |
Flow Rate:
|
0.1 CFM (2.83 LPM) |
Laser Source:
|
Laser Diode |
Zero Count
Level:
|
<1 count / 5 minutes (per JIS)
|
Calibration:
|
NIST Traceable |
Count Modes:
|
Automatic, Manual, Beep, Concentration,
Cumulative/Differential |
Data Storage:
|
3,000 sample records, rotating
buffer (includes particle and environmental
data, plus location and time) |
Communication
Modes:
|
RS-232 via RJ-45 to
PC
RS-485 MODBUS |
Supporting
Software:
|
LMS XChange, LMS Express, LMS
Express RT and LMS Express RT+,
LMS Net |
| Environmental Sensors:
|
Optional (up to 4): Temperature,
Relative Humidity, Air Velocity,
Differential Pressure or others
|
| Concentration Limits:
|
4,000,000 particles / ft3
@ 5% Coincidence Loss |
Touch Screen
Display:
|
5.7” (14.47 cm) Color, 320x240 |
| Printer: |
Built-in Thermal Printer |
| Reports: |
FS-209E, ISO-14644-1, BS-5295
& EC GMP |
| Alarm: |
Internal, Adjustable Alarm Buzzer,
External Alarm Connector. Alarms
on Counts, Low Battery, Sensor Failure |
| Enclosure: |
Stainless steel |
| Sample Output: |
Internally filtered to HEPA standards
(>99.997% @ 0.3 µm) |
Vacuum
Source:
|
Internal Pump, Automatic Flow
Control |
| Power: |
85-264 VAC, 47-63 Hz |
Battery:
|
Li-Ion, Removable & Rechargeable |
Dimensions:
|
8” (l) x 8.5” (w) x 8.9” (h) [20.32
x 21.59 x 22.60 cm] |
| Weight:
|
With Battery: 14.75
lbs (6.7 kg) Without Battery:
12.75 lbs (5.8 kg) |
| Environmental
Conditions |
| Operating:
|
50°F to 104°F (10°C
to 40°C) / 20% to 95% non-condensing
|
| Storage:
|
14°F to 122°F (-10°C
to 50°C) / Up to 98% non-condensing
|
| Accessories |
| Included:
|
Operating Manual on
CD, Isokinetic Sample Probe with
Tripod and Tubing, Purge Filter,
Removable/Rechargeable Li-Ion Battery,
Power Cable and Spare Fuse, LMS
XChange Software, Printer, Printer
Paper |
| Optional:
|
Shipping Case with
Wheels, Cleanroom Cart, Temperature/Relative
Humidity Probe, Air Velocity Probe,
Differential Pressure Probe, Filter
Scanning Probe, High Pressure Diffuser,
Spare Removable/Rechargeable Li-Ion
Battery, External Battery Charger,
LMS Express Software, LMS Express
RT Software, LMS Express Software
RT+, Validation Documentation |
|
|
| |
| [ Back to