| Specifications |
Size Range:
|
0.2 - 2μm |
Channel
Sizes:
|
Standard: 0.2, 0.3μm
Standard: 0.2, 0.5μm
Optional: 0.2, 0.3, 0.5, 1.0μm
Optional: 0.2, 0.3, 0.5, 2.0μm |
Counting
Efficiency:
|
50% @ 0.2μm; 100% for particles
> 0.3μm (per JIS) |
Flow Rate:
|
0.1 CFM (2.83 LPM) |
Laser Source:
|
Laser Diode |
Zero Count
Level:
|
<1 count / 5 minutes (per JIS) |
Concentration
Limit :
|
2,000,000 Particles / ft3 @ 5%
Coincidence Loss |
Calibration:
|
NIST Traceable |
Vacuum
Requirements:
|
External Vacuum >18” (45.72)
of Hg |
Air Flow:
|
Internally Monitored |
Communication
Modes:
|
RS-485 MODBUS |
LED Indicators:
|
Power, Flow, Service, Sampling |
Data Storage:
|
250 Samples |
Connector:
|
RJ45 |
Supporting
Software:
|
LMS XChange, Lighthouse Monitoring
System; LMS Express; Express RT
and RT+; LMSNet |
| Environmental
Sensors: |
(Optional) Temperature/Relative
Humidity Probe 32 - 122F (0-50C)+/-1F
(0.5C); 0 - 100% RH +/-2% |
External
Alarm Output:
|
Normally Open Dry Contact Rated
0-60 V AC/DC 1 Amp |
Enclosure:
|
Stainless steel |
| Power: |
24 VDC |
Dimensions:
|
1.7”(l) x 4.2”(w) x 3”(h) [4.3
x 10.7 x 7.6 cm] |
| Weight:
|
0.97 lb (0.44 kg) |
| Environmental
Conditions |
| Operating:
|
50°F to 104°F (10°C
to 40°C) / 20% to 95% non-condensing |
| Storage:
|
14°F to 122°F (-10°C
to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included:
|
Operating Manual on
CD |
| Optional:
|
Printed Operating Manual;
RS-485 Hub with Power Supply; RS-232
Communication Cable and Adapter;
USB to Serial Port Converter; Network
Adapter; Wireless Network Adapter;
Isokinetic Sampling Probes; Vacuum
Tubing; Sample Tubing; Data Cable;
Purge Filter; Power Supply; Temperature/
Relative Humidity Probe; 8’ T/RH
Probe Extension Cable; REMOTE Mounting
Bracket; LMS XChange, LMS Express;
Express RT and RT+, LMSNet software;
RS-485 Communication Kit. |
| Sample System Integration |
![]() |