Airborne Particle Counters
Remote Products > 5014P
REMOTE 5014P

Incorporating the latest in laser optical particle counting technology, Lighthouse designed the REMOTE 5014P airborne particle counter for continuous troublefree operation.

With sensitivity of 0.5 at a flow rate of 0.1 CFM (2.83 LPM), the REMOTE 5014P provides real-time continuous data collection at a cost-effective price per point. Using an internal vacuum source, the compact REMOTE 5014P particle counter can be installed where space is at a premium.

The REMOTE 5014P integrate seamlessly into large facility monitoring and management systems and transfer up to 6 channels of simultaneous particle count data using RS-485 Modbus.

Designed and built by Lighthouse, a name you can trust.



Virtual Demo ]
 
Internal Vacuum Pump Direct Mount Isokinetic Probe RJ45 Connector LED Status Alarm Relay Four 4-20mA Inputs


Features Benefits
  • 0.5 - 25.0μm Range
  • 0.1 CFM (2.83 LPM) Flow Rate with Internal Pump
  • Up to 6 Channels of Simultaneous Data
  • All Models Meet JIS Standards
  • Compact Size
  • Stainless Steel Enclosure
  • Up to Four 4-20 mA Inputs (30 mW)
  • LED Status Indicators
  • External Alarm Connector
  • RS-485 Modbus
  • Facility Monitoring System Interface
  • Flow-controlled Internal Vacuum Pump
  • Designed for Reliability
  • 2 Year Warranty
  • Installs in Limited Space
  • Cost Effective Price per Point
  • Proven Data Integrity
  • Proven Reliability
  • Continuous Real-Time Data
  • Easy System Integration
  • International Support
  • Low Cost of Ownership
 
 
Specifications
Size Range:
0.5 - 25.0 μm
Channel Sizes:
Standard: 0.5, 1.0, 5.0, 10.0 μm
Counting Efficiency:
50% @ 0.5 μm; 100% for particles > 0.75 μm (per JIS)
Flow Rate:
0.1 CFM (2.83 LPM)
Laser Source:
Laser Diode
Zero Count Level:
<1 count / 5 minutes (per JIS)
Concentration Limits:
2,000,000 Particles / ft3 @ 5% Coincidence Loss
Calibration:
NIST Traceable
Vacuum Requirements:
Internal Vacuum Pump
Air Flow:
Internally Monitored and Controlled
Communication Modes:
RS-485 MODBUS
LED Indicators:
Power, Flow, Service, Sampling
Data Storage:
3,000 Samples
Connectors:
RJ45
Supporting Software:
LMS XChange; Lighthouse Monitoring System; LMS Express, Express RT and RT+; LMSNet
Up to Four 4-20 mA Inputs: Can be used with the Lighthouse standard 4-20 mA Temperature/Relative Humidity, Air Velocity or Differential Pressure Probes or any other 4-20 mA device
External Alarm Output:
Normally Open Dry Contact Rated 0-60 V AC/DC 1 Amp
Enclosure:
Stainless Steel
Power: 24 VDC
Dimensions:
9”(l) x 3.5”(w) x 6”(h) [22.86 x 8.89 x 15.24 cm]
Weight: 3.19 lbs (1.45 kg)
Environmental Conditions
Operating: 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage: 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing
Accessories
Included: Operating Manual on CD; LMS XChange; Power Supply; Power Cord
Optional: Printed Operating Manual; REMOTE Display; REMOTE Display 10-foot Extension Cable; Network Adapter; Wireless Network Adapter; Isokinetic Sampling Probe; Sample Tubing; Data Cabling; Purge Filter; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; RS-232 Interface Cable and Adapter; LMS Express, Express RT, RT+ and LMSNet software
Sample System Integration
 
 