| Specifications |
Size Range:
|
0.5 - 25.0 µm |
Channel
Sizes:
(Specify During Order)
|
Standard: 0.5, 5.0μm;
Optional: 0.5, 10.0; 0.5, 25.0 μm |
Counting
Efficiency:
|
50% @ 0.5μm; 100% for particles
> 0.75μm (per JIS) |
Flow Rate:
|
1.0 CFM (28.3 LPM) |
Laser Source:
|
Laser Diode |
Zero Count
Level:
|
<1 count / 5 minutes (per JIS) |
Concentration
Limits:
|
500,000 Particles/ft3 @ 5% Coincidence
Loss |
Calibration:
|
NIST Traceable per ISO 21501-4 |
Vacuum
Requirements:
|
External Vacuum >18" (45.72)
of Hg |
Air Flow:
|
Critical Orifice controlled |
Communication
Modes:
|
4-20mA |
LED Indicators:
|
Power, Service |
Connector:
|
Industrial RJ45 |
Supporting
Software:
|
Lighthouse Monitoring System,
LMS Express, LMS Express RT, LMS
Express RT Plus, LMS Net |
Enclosure:
|
316L Stainless Steel, Wash Down
Resistant |
| Power: |
24 VDC |
Dimensions:
|
5.4” (l) x 4.6” (h) x 1.65” (d)
[13.72 x 11.68 x 4.19 cm] |
| Weight:
|
17.3 oz (0.49 kg) |
| Environmental
Conditions |
| Operating:
|
50°F to 104°F (10°C
to 40°C) / 20% to 95% non-condensing
|
| Storage:
|
14°F to 122°F (-10°C
to 50°C) / Up to 98% non-condensing
|
| Accessories |
| Included:
|
Operating Manual on
CD |
| Optional:
|
Isokinetic Sampling
Probe, Sample Tubing, Cabling, Purge
Filter, RS-485 Hub |