| Airborne Particle Counters |
| Remote Products
> 5104V |
| REMOTE 5104V |
| Compatible
with VHP, the REMOTE 5104V
meets the stringent requirements
of today’s aseptic processing
applications. In addition, the
REMOTE 5104V
was designed to operate in CLASS
1 DIV 2 hazardous environments
found in many challenging applications.
The REMOTE 5104V
uses a critical orifice and external
vacuum source for flow control.
Built-in flow monitoring assures
accuracy and reliability of the
sensor’s performance.
With a sensitivity of 0.5 micron
and a monitored flow rate of 1.0
CFM (28.3 LPM), the REMOTE
5104V provides real-time
continuous data collection at
a cost-effective price per point.
The REMOTE 5104V integrates
seamlessly with large facility
monitoring and building management
systems. It transfers up to 4
channels of simultaneous particle
count data using RS-485 Modbus.
Designed and built by Lighthouse
Worldwide Solutions,
a name you can trust. |

[ Virtual
Demo ] |
|
|
| Features |
Benefits |
- VHP Compatible
- Class 1 Division 2 Rating
- Built-in Flow Monitoring
- 0.5 - 5.0 μm (optional
channels up to 25 μm)
- 1.0 CFM (28.3 LPM) Flow Rate
- Up to 4 Channels of Simultaneous
Data
- Meets or Exceeds JIS Standards
- Real-time Continuous Monitoring
- Small Size
- 316L Stainless Steel, Wash
Down Resistant/Splash-Tight
Enclosure
- Status Indicators
- RS-485 Modbus
- Interfaces to Facility Monitoring
Systems
- Uses External Vacuum Source
- Designed for Reliability
- 2 Year Warranty
- International Support
|
- Versatile Mounting Options
- Easy to Install when Space
is an Issue
- Cost Effective Price per Point
- Proven Data Integrity
- Simple Integration into FMS
Systems
- Low Cost of Ownership
|
| Applications |
Industries |
- Glass Coating
- Aseptic Processing
- Isolators
- Cleanrooms
|
- Aerospace Industry
- Pharmaceutical Industry
- Biotechnology
- Medical Applications
|
|
|
| |
| Specifications |
Size Range:
|
0.5 - 25.0 µm |
Channel
Sizes:
|
Standard: 0.5, 5.0 µm
Optional 4 Channels: 0.5, 1.0, 5.0,
25.0 µm |
Counting
Efficiency:
|
50% @ 0.5μm; 100% for particles
> 0.75μm (per JIS) |
Flow Rate:
|
1.0 CFM (28.3 LPM) |
Laser Source:
|
Laser Diode |
Zero Count
Level:
|
<1 count / 5 minutes (per JIS) |
Concentration
Limits:
|
500,000 Particles/ft3 @ 5% Coincidence
Loss |
Calibration:
|
NIST Traceable |
Vacuum
Requirements:
|
External Vacuum >18" (45.72)
of Hg |
Air Flow:
|
Internally Monitored |
Communication
Modes:
|
RS-485/MODBUS |
LED Indicators:
|
Power, Flow, Service, Sampling |
Connector:
|
Industrial RJ45 |
Supporting
Software:
|
Lighthouse Monitoring System,
LMS Express, LMS Express RT, LMS
Express RT Plus, LMS Net |
Enclosure:
|
316L Stainless Steel, Wash Down
Resistant/Splash-Tight |
| Power: |
24 VDC |
Dimensions:
|
4.2” (l) x 2.0” (w) x 3.7” (h)
[10.7 x 5.08 x 9.398 cm] |
| Weight:
|
15.6 oz (0.44 kg) |
| Environmental
Conditions |
| Operating:
|
50°F to 104°F (10°C
to 40°C) / 20% to 95% non-condensing
|
| Storage:
|
14°F to 122°F (-10°C
to 50°C) / Up to 98% non-condensing
|
| Accessories |
| Included:
|
Operating Manual on
CD |
| Optional:
|
Isokinetic Sampling
Probe, Sample Tubing, Cabling, Purge
Filter, Power Supply |
 |
|
|
| |
| [ Back to top ] |
|
|
|
|
| |
| |
|
| |
| |
|